- Measures Total Cyanide by UV- digestion with amperometric detection technique according to ASTM Method D7511-09 and OIA 1678.
- Technique has fewer analytical interferences (positive and negative) and consequently supports a lower method detection limit (MDL) and a broader measurement range.
- Eliminates preliminary 2-hour acid distillation step prior to analysis.
- Amperometric detection eliminates use of hazardous pyridine, barbutaric acid reagents required in cyanide analysis by spectrophotometric method.
- A fully automated system with a industry accredited Program Logic controller (PLC) and Human Machine Interface (HMI)
- Built in diagnostic features for reagent level alarm and the component malfunctions
- Data communication compatible with required SPCB/CPCB norms
- Operating principle: UV digestion with amperometry
- Measurement Technique: Amperometric detection with three electrode system
- Measuring Range: 10 – 2000 ppb/ 2000-20000 ppb
- Reference Method: ASTM 7511-09e2
- Measurement Accuracy: ±20 ppb for 10-200 ppb
- ±10% of reading for ≥200 ppb
- Calibration: Single point calibration
- Sample injection: Automatic
- Sampling interval: 30 min
- Analysis time: 30 min
- Operating temperature: 20oC to 50oC
- Display: HMI full touch screen
- Reagents: NaOH, H2SO4, distilled water, Hypophosphorous acid, ascorbic acid, Bismuth nitrate
- Power: 230 V AC/ 60 Hz
- Instrument dimensions:1200 X 400 X 600 mm
- Instrument weight: 70kg