The OPTITEMP TT 51 R is a universal, programmable high-performance 2-wire rail-mounted temperature transmitter with HART® communication. The galvanically isolated dual input device works with RTD elements, thermocouple (TC) elements or other inputs (mV and Ω). The device is available in 2-, 3- or 4-wire circuit. It also features the option of a 4-wire circuit for two Pt100 inputs. The transmitter stands for proven stability, reliability as well as high accuracy and boasts a minimum measuring error of ±0.05% of span in 5 years. The transmitter offers extended diagnostic information (e.g. open or short circuit monitoring etc.). Its unique "SmartSense" function allows insulation resistance monitoring to prevent measuring errors.
Show more
The OPTITEMP TT 51 R is SIL2 approved (IEC 61508) and thus the ideal solution for use in Safety Instrumented Systems (SIS). The temperature transmitter is also available as intrinsically safe. As an added bonus, the ConSoftTM configuration software makes the transmitter start-up particularly simple. The temperature transmitter is designed to be mounted on a DIN rail. A dedicated head-mounted version is available with the OPTITEMP TT 51 C.
Product highlights
- Highly precise, completely universal digital device, with dual sensor input
- Configuration, installation and maintenance via software ConSoft, HART® protocol or EDD
- Optional redundant 4-wire circuit for 2 x Pt100
- Best in class for long-term stability: ±0.05% of span in 5 years
- Fast response time, high speed update (300 ms)
- High reliability due to 50-point linearisation, sensor drift detection and sensor backup
- Sensor break monitoring, failure detection and error correction
- Isolation resistance monitoring (SmartSense): fail-safe up to 95% relative air humidity
- Rugged design: vibration-resistant up to 10 g
- Galvanic isolation: 1500 VAC
- Conforms to NAMUR in accordance with NE 43, NE 53, NE 89, NE 107
- Conforms to EMC in accordance with EN 61326 and NE 21
- SIL2 certified as per IEC 61508
- Can be configured via ConSoft while staying connected with a sensor installed in hazardous areas
- Designed for installation on a rail (in accordance with DIN EN 60715)
- Available as intrinsically safe: ATEX, IECEx (zone 0, 1, 2)
Technical data
Transmitter mounting | Rail-mounted |
Transmitter channels MAX | 2 |
Galvanic isolation | Yes |
Galvanic isolation (VAC) | 1500 VAC |
Transmitter input | Potentiometer (Ω), Resistance (RTD), Thermocouple (TC), Voltage (mV) |
RTD input | Cu10, Ni100, Ni1000, Ni120, Pt10, Pt100, Pt1000, Pt200, Pt50, Pt500 |
RTD input / wiring | 2-wire, 3-wire, 4-wire |
TC Input | B, C, D, E, J, K, N, R, S, T |
Potentiometer input (Ω) | 100…4000 Ω |
Voltage input (mV) | -10…+1000 mV |
Transmitter programmable | Yes |
Configuration method | HART handheld, PC/ConSoft software |
Power supply (VDC) | 10…36 VDC 10…30 VDC (Ex) |
Ambient temperature | -20…70°C/ -4…158°F |
Diagnostic functions | Isolation monitoring (SmartSense), Sensor backup function, Sensor break detection, Sensor drift detection, Sensor error correction, Sensor short circuit detection, Signal filtering function |
Approvals/Standards
Certificates/Approvals Ex | ATEX, IECEx |
Certificates/Approvals Safety | SIL2 |
NAMUR compliance | NE 107, NE 21, NE 43, NE 53, NE 89 |
Communication
Digital outputs | HART® |
Analogue outputs | 4…20 mA |